2D ultrashort echo‐time functional lung imaging

Loading...
Thumbnail Image

Date

2020-07-11

Authors

Balasch, Anke

Metze, Patrick

Stumpf, Kilian

Beer, Meinrad

Büttner, Susanne M.

Rottbauer, Wolfgang

Speidel, Tobias

Rasche, Volker

Journal Title

Journal ISSN

Volume Title

Document version

publishedVersion

Distributor

Universität Ulm

Published in

Journal of Magnetic Resonance Imaging, 2020

Abstract

Background Imaging of the lung by MRI is challenging due to the intrinsic low proton density and rapid T2* relaxation. MRI methods providing lung parenchyma and function are in demand. Purpose To investigate the feasibility of two‐dimensional ultrashort echo‐time (2D UTE) imaging for lung function assessment. Study Type Prospective. Population Eleven healthy volunteers. Field Strength/Sequence 3T, 2D tiny golden angle UTE (2D‐tyUTE). Assessment The applicability of breath‐hold (BH) and self‐gated (SG) 2D‐tyUTE for quantification of the lung parenchyma signal‐to‐noise ratio (SNR), proton fraction (fP), fractional ventilation (FV), and perfusion (f) was investigated. Dependencies on repetition time (BHS/I1/I2) and respiratory phase (expiration [EX], inspiration [IN]) were investigated and compared between smokers and nonsmokers. Statistical Tests Analysis of variance (ANOVA), Kendell's W. Results Significant differences of SNR (EX: 10.98 ± 3.19(BHS), 14.58 ± 3.89(BHI1), 17.59 ± 4.92(BHI2), 11.00 ± 5.42(SG); IN: 7.17 ± 2.07(BHS), 9.51 ± 2.37(BHI1), 10.49 ± 2.33(BHI2), 10.00 ± 4.14(SG)) (P < 0.05 for all cases) were observed between the different approaches. Where fP in expiration (0.41 ± 0.13) was independent of the BH imaging technique, it was slightly higher in SG (0.44 ± 0.06). FV was reproducible among the BH techniques (0.41 ± 0.10), but significantly lower in SG (0.21 ± 0.06) (P < 0.05). A moderate correlation (R2 = 0.47, P < 0.01) was observed between the breathing amplitude and FV. No significant differences between BH and SG were observed for the perfusion analysis (EX: 3.50 ± 1.29 mL/min/mL [BHS]; IN: 2.36 ± 1.05 mL/min/mL [BHS]). Significant differences in fP were found between smokers (0.48 ± 0.11 BH) and nonsmokers (0.37 ± 0.12 BH) in expiration. Data Conclusion This study demonstrates the feasibility of 2D‐tyUTE for successful quantification of relevant lung function parameters at 3T within clinically attractive acquisition times. The low spatial resolution into the slice selection direction may limit the final sensitivity and needs further clinical evaluation. Level of Evidence 2 Technical Efficacy Stage 1 J. MAGN. RESON. IMAGING 2020;52:1637–1644.

Description

Faculties

Medizinische Fakultät

DFG Project uulm

EU Project uulm

Other projects uulm

Citation

Is version of

Has version

Supplement to

Has erratum

Erratum to

Has Part

Part of

DOI external

DOI external

10.1002/jmri.27269

Institutions

Periodical

Degree Program

DFG Project THU

EU Project THU

Other projects THU

Series

Collections

Endorsement

Review

Supplemented By

Referenced By

Zeitraum:

Creative Commons license

Except where otherwised noted, this item's license is described as CC BY-NC 4.0 International