Genauigkeit von Messungen an Implantaten und anatomischen Strukturen in der zahnärztlichen Radiologie

dc.contributor.authorWerner, Patrick
dc.date.accessioned2016-03-15T10:42:12Z
dc.date.available2016-03-15T10:42:12Z
dc.date.created2014
dc.description.abstractLength measurements on radiographs take an important role in dental practice. Objectives of this study were, first, the comparison of the X-ray dental film method, the panoramic radiography and the Cone beam computed tomography (CBCT) and secondly, the comparison of two panoramic dental x-ray machines and three different CBCT-units in terms of their metric accuracy. In this study, five mandibular specimens by human body donors were used. Implants were placed in three of the five mandibles. For the experimental procedure dental films, panoramic radiographs and CBCT-reconstructions were made. In each radiograph the length of a defined measuring distance was compared with the measurement of the same distance on the body donor. The examples in this investigation have demonstrated that accurate measurements on radiographs or in three-dimensional-reconstructions are possible. There were cases where the lengths measured in the X-ray images were equal to the measurements at the specimens. The CBCT turned out to be a very accurate measuring instrument. By comparing the three CBCT-units no difference appeared concerning the metric measuring accuracy. But in our case the dental film and the panoramic radiography have proved that they also can provide very accurate results. However, the dental film and the panoramic radiography are also more prone to measurement errors and misinterpretation. The transmission of radiographic measurement result in the clinical situation is a major problem. The CBCT has emerged as a very accurate measuring instrument. As a standard tool for basic examination or treatment in simple cases, however, the panoramic radiography and the dental film are sufficient and recommended. In more complex cases modern three-dimensional radiological diagnostics and planning should be used.
dc.identifier.doihttp://dx.doi.org/10.18725/OPARU-3605
dc.identifier.ppn826705979
dc.identifier.urlhttps://oparu.uni-ulm.de/xmlui/123456789/3632
dc.identifier.urnhttp://nbn-resolving.de/urn:nbn:de:bsz:289-vts-95152
dc.language.isode
dc.publisherUniversität Ulm
dc.rightsStandard (ohne Print-On-Demand)
dc.rights.urihttps://oparu.uni-ulm.de/xmlui/license_opod_v1
dc.subjectCBCT
dc.subjectDigitale Volumentomographie
dc.subjectDVT
dc.subjectPanoramaschichtaufnahme
dc.subjectZahnfilm
dc.subject.ddcDDC 610 / Medicine & health
dc.subject.gndImplantat
dc.subject.gndLängenmessung
dc.subject.gndOrthopantomogramm
dc.subject.gndRöntgenbild
dc.subject.gndVolumentomographie
dc.subject.meshRadiography, dental
dc.titleGenauigkeit von Messungen an Implantaten und anatomischen Strukturen in der zahnärztlichen Radiologie
dc.typeDissertation
uulm.affiliationGeneralMedizinische Fakultät
uulm.bibliographieuulm
uulm.categoryPublikationen
uulm.freischaltungVTS2015-05-12T14:23:58Z
uulm.peerReviewnein
uulm.shelfmarkW: W-H 14.183
uulm.typeDCMIText
uulm.updateStatusURNurl_update_general
uulm.vtsID9515

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
vts_9515_14374.pdf
Size:
2.9 MB
Format:
Adobe Portable Document Format
Description:
125 S.

Collections