Abstract

Postoperative implant-associated infections are a severe complication in orthopaedics and trauma surgery. To address this problem, a novel implant coating was recently developed, which allows for the release of low concentrations of bactericidal silver. For an intended use on load-bearing endoprostheses, stable bone integration is required. The aim of the present study was to evaluate the biocompatibility and osseointegration of titanium implants with the novel coating in a mechanically loaded bone-defect model in sheep.Silver-coated devices were implanted into weight-bearing and non-weight-bearing tibial and femoral bonedefects whereas, in the control group, uncoated titanium implants were inserted. The bony integration of the implants was assessed mechanically and histologically after 6 months. Silver concentrations were assessed in peripheral blood, liver, kidney and local draining lymph nodes as well as at the implantation site. After 6 months, shear strength at the interface and bone apposition to the implant surface were not significantly different between coated and uncoated devices. Mechanical loading reduced bony integration independently of the coating. Silver content at the implantation site was larger in the group with silver-coated implants, yet it remained below toxic levels and no cytotoxic side effects were observed. Concluding, the novel antibacterial silver coating did not negatively influence bone regeneration or implant integration under mechanically unloaded and even loaded conditions, suggesting that the silver coating might be suitable for orthopaedic load-bearing implants, including endoprostheses.

Description

Faculties

DFG Project uulm

EU Project uulm

Other projects uulm

Citation

Is version of

Has version

Supplement to

Supplemented by

Has erratum

Erratum to

Has Part

Part of

DOI external

DOI external

10.22203/eCM.v039a16

Institutions

Periodical

Degree Program

DFG Project THU

EU Project THU

Other projects THU

Series

Collections

Endorsement

Review

Supplemented By

Referenced By

Zeitraum:

Creative Commons license

Except where otherwised noted, this item's license is described as CC BY-SA 4.0 International