<link rel="stylesheet" href="styles.cf390c6bfbe7555f.css">

Biomechanics of a calcar loading and a shortened tapered femoral stem: comparative in‑vitro testing of primary stability and strain distribution

dc.contributor.authorFreitag, Tobias
dc.contributor.authorBieger, Ralf
dc.contributor.authorKiefer, Hartmuth
dc.contributor.authorDornacher, Daniel
dc.contributor.authorReichel, Heiko
dc.contributor.authorIgnatius, Anita
dc.contributor.authorDürselen, Lutz
dc.date.accessioned2021-12-09T15:18:34Z
dc.date.available2021-12-09T15:18:34Z
dc.date.issued2021-09-07de
dc.description.abstractPurpose The most common femoral short stems available on the market can, in principle, be divided with regard to their anchoring concepts into a calcar loading and a shortened tapered design. The purpose of this study was to compare the primary stability and stress-shielding of two short stems, which correspond to these two different anchoring concepts. Methods Using seven paired fresh frozen human cadaver femurs, primary axial and rotational stabilities under dynamic load (100–1600 N) were evaluated by miniature displacement transducers after 100,000 load cycles. Changes in cortical strains were measured before and after implantation of both stem types to detect implant-specific load transmission and possible stress-shielding effects. Results Reversible and irreversible micromotions under dynamic load displayed no significant differences between the two implants. Implantation of either stem types resulted in a reduction of cortical strains in the proximal femur, which was less pronounced for the calcar loading implant. Conclusions Both short stems displayed comparable micromotions far below the critical threshold above which osseointegration may disturbed. Neither short stem could avoid proximal stress-shielding. This effect was less pronounced for the calcar loading short stem, which corresponds to a more physiological load transmission.de
dc.description.versionpublishedVersionde
dc.identifier.doihttp://dx.doi.org/10.18725/OPARU-40297
dc.identifier.urlhttps://oparu.uni-ulm.de/xmlui/123456789/40373
dc.identifier.urnhttp://nbn-resolving.de/urn:nbn:de:bsz:289-oparu-40373-4
dc.language.isoende
dc.publisherUniversität Ulm
dc.relation1.doi10.1186/s40634-021-00388-1
dc.rightsCC BY 4.0 Internationalde
dc.rights.urihttps://creativecommons.org/licenses/by/4.0/
dc.subjectShort stemde
dc.subjectPrimary stabilityde
dc.subjectStress-shieldingde
dc.subjectTotal hip arthroplastyde
dc.subjectMicromotionsde
dc.subjectMechanical stressde
dc.subjectMetaphysealde
dc.subject.ddcDDC 570 / Life sciencesde
dc.subject.gndLeichede
dc.subject.lcshTotal hip replacementde
dc.subject.meshArthroplasty, Replacement, Hipde
dc.subject.meshCadaverde
dc.subject.meshStress, Mechanicalde
dc.titleBiomechanics of a calcar loading and a shortened tapered femoral stem: comparative in‑vitro testing of primary stability and strain distributionde
dc.typeWissenschaftlicher Artikelde
source.articleNumber74de
source.identifier.eissn2197-1153de
source.publisherSpringer Opende
source.titleJournal of Experimental Orthopaedicsde
source.volume8de
source.year2021de
uulm.OAcosts1608,88 / 2021
uulm.OAfundingOpen-Access-Förderung durch die Medizinische Fakultät der Universität Ulm
uulm.affiliationSpecificUKU. Klinik für Orthopädiede
uulm.affiliationSpecificZentrum für Traumaforschung der Universität Ulm
uulm.affiliationSpecificUKU. Institut für Unfallchirurgische Forschung und Biomechanikde
uulm.bibliographieuulm
uulm.categoryPublikationen
uulm.cooperationLukas-Krankenhaus gGmbHde
uulm.peerReviewjade
uulm.typeDCMITextde
uulm.updateStatusURNurn_new

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Freitag_2021.pdf
Size:
1.55 MB
Format:
Adobe Portable Document Format
Description:

Collections