The effect of composite attachments on the force and moment systems generated by thermoplastic PET-G aligners during the rotation of a mandibular canine : an in-vitro study
Loading...
Date
2019-09-10
Authors
Mikhaiel, Bishoy
Journal Title
Journal ISSN
Volume Title
Publication type
Document version
Distributor
Universität Ulm
Published in
Abstract
INTRODUCTION:
The influence of different composite attachments on the three dimensional force-moment (F/M) components exerted by aligners of different thicknesses during laboratory simulation of mesio- and distorotation of a lower canine was examined.
MATERIALS AND METHODS:
An acrylic mandibular model with a separated right canine mounted on a hexapod via a 3D F/M sensor was used. Duran+® aligners (Scheu Dental, Germany) of thickness 0.5, 0.625, and 0.75 mm were fabricated on plaster models with the measurement tooth in its neutral position. Three different composite attachment variations were designed on plaster casts and attached to the tooth using a corresponding template aligner. The F/M values were recorded during stepwise mesio- and distorotation of tooth 43 in 1° steps up to ±15°, corresponding to 0.5 mm displacements of the tooth's proximal contacts and the rotational stiffness of each attachment variation was calculated. Each rotation step included removal and reseating of the aligner on the acrylic model. Three aligners of each thickness were produced for each direction of rotation. Each aligner underwent three trial repetitions.
RESULTS:
The median rotational moments for the 0.5 mm aligner and 15° distorotation of tooth 43 when not applying an attachment was 27.49 Nmm (interquartile range, IQR 1.45 Nmm). The corresponding values for the 0.625 and 0.75 mm aligners were 41.04 Nmm (IQR 5.62 Nmm) and 42.48 Nmm (IQR 2.17 Nmm), respectively. The average rotational moments for distorotation were 15% higher than for mesiorotation (p = 0.01). Applying Canine Rotation Attachment (CRA) during distorotation and Ellipsoid attachment during mesiorotation demonstrated highest average rotational stiffnesses (9.33 Nmm/° and 11.13 Nmm/°, respectively). Relatively high collateral F/M components, namely an intrusive force besides labiolingual and mesiodistal tipping moments varying with each attachment variation were observed.
CONCLUSION:
To avoid overloading of periodontal structures, derotation of lower canines should be limited to 10° per setup step, leading to rotational moments of about 15 Nmm. Due to their similarity in biomechanical behaviour with 0.75 mm aligners, 0.625 aligners may be omitted from the aligner sequence. Applying the tested attachments increases the resulting rotational moments. Further studies are required to investigate additional attachment modifications and geometries on minimizing collateral F/M components during canine derotation.
Description
Faculties
Medizinische Fakultät
DFG Project uulm
EU Project uulm
Other projects uulm
Citation
Is version of
Has version
Supplement to
Supplemented by
Has erratum
Erratum to
Has Part
Part of
DOI external
DOI external
Institutions
Periodical
Degree Program
DFG Project THU
EU Project THU
Other projects THU
Series
Collections
Endorsement
Review
Supplemented By
Referenced By
Zeitraum:
Creative Commons license
Except where otherwised noted, this item's license is described as Standard