Evaluierung von hochkonzentriertem jodhaltigem Kontrastmittel in der Multidetektor-Computer-Tomographie der Aorta mit automatischem Bolus-Tracking
Loading...
Date
Authors
Friedrich, Patricia Claude
Journal Title
Journal ISSN
Volume Title
Publication type
Document version
Distributor
Universität Ulm
Published in
Abstract
Purpose: To objectively evaluate the benefit of a contrast medium with high concentration Iodine for Multi-Detector CT Angiography (MDCTA) of the aorta using automatic bolus tracking. -
Materials and methods: A total of 100 consecutive MDCTA examinations of the entire aorta were evaluated. All scans were performed on a 40 channel MDCT scanner. A contrast medium with a high Iodine concentration of 400 mg/ml (Imeron 400, Bracco, Milano, Italy) was administered by power injector at a dose of 1.2 ml/kg BW and at a rate of 3.5 ml/sec, followed by a saline chaser. Scanning was initiated by automated bolus tracking with a delay of 6 sec after attaining a measured threshold of 150 HU in the descending aorta. Time-density distributions calculated from the ascending aorta downstream to the common femoral arteries were graded as from poor to excellent based on contrast density, homogeneity, and timing of the scan acquisition. -
Results: Evaluated aortic lesions included 23 thoracic and 39 abdominal aneurysms and 30 aortic dissections. The mean (± SD) overall arterial density of MDCTA was 306 ± 66 HU (range: 129 - 476 HU). Averaged arterial contrast density was rated good or better in 92 cases (> 200 HU) while contrast homogeneity was rated good or better in 83 cases. Automatic bolus tracking permitted accurate scanning during peak contrast enhancement in 87 cases. The 13 mistimed examinations comprised 12 MDCTA studies that were started early and one that was started late. In all early scans the contrast density was still considered good or better. -
Conclusion: The combination of high concentration contrast medium and automatic bolus tracking allows reliable MDCTA of the entire aorta including the pelvic axis with excellent and homogeneous contrast distribution in a variety of aortic pathologies.
Description
Faculties
Medizinische Fakultät
Institutions
DFG Project uulm
EU Project uulm
Other projects uulm
Citation
Is version of
Has version
Supplement to
Supplemented by
Has erratum
Erratum to
Has Part
Part of
DOI external
DOI external
Institutions
Periodical
Degree Program
DFG Project THU
EU Project THU
Other projects THU
Series
Collections
Endorsement
Review
Supplemented By
Referenced By
Zeitraum:
Creative Commons license
Except where otherwised noted, this item's license is described as Standard (Fassung vom 03.05.2003)