Klinische Genauigkeit der schablonengestützten Implantatinsertion

Erstveröffentlichung
2016-06-07Autoren
Naziri, Eleni
Gutachter
Schramm, AlexanderLuthardt, Ralph
Dissertation
Fakultäten
Medizinische FakultätInstitutionen
Bundeswehrkrankenhaus Ulm (BWK)UKU. Klinik für Mund-, Kiefer- und Gesichtschirurgie
UKU. Klinik für Zahnärztliche Prothetik
Zusammenfassung
Abstract
Objectives: The purpose of this study was to assess the accuracy of computer-assisted implant insertion based on computed tomography and template-guided implant placement.
Material and methods: A total of 246 implants were placed with the aid of 3D-based transfer templates in 181 consecutive partially edentulous patients. Five groups were formed on the basis of different implant systems, surgical protocols and guide sleeves. After virtual implant planning with the CoDiagnostiX Software, surgical guides were fabricated in a dental laboratory. After implant insertion, the actual implant position was registered intraoperatively and transferred to a model cast. Deviations between the preoperative plan and postoperative implant position were measured in a follow-up computed tomography of the patient’s model casts and image fusion with the preoperative computed tomography.
Results: The median deviation between preoperative plan and postoperative implant position was 1.0 mm at the implant shoulder and 1.4 mm at the implant apex. The median angular deviation was 3.6º. There were significantly smaller angular deviations (P=0.000) and significantly lower deviations at the apex (P=0.008) in implants placed for a single-tooth restoration than in those placed at a free-end dental arch. The location of the implant, whether in the upper or lower jaw, did not significantly affect deviations. Increasing implant length had a significant negative influence on deviations from the planned implant position. There was only one significant difference between two out of the five implant systems used.
Conclusion: The data of this clinical study demonstrate the accuracy and predictable implant placement when using laboratory-fabricated surgical guides based on computed tomography.
Keywords: dental implant planning, surgical guides, surgical templates, computer-assisted surgery, computer-aided surgery, guided implant surgery
Erstellung / Fertigstellung
2015
Schlagwörter
[GND]: Zahnimplantat | Computerassistierte Chirurgie[MeSH]: Dental implantation; Methods | Surgery, computer-assisted
[Freie Schlagwörter]: 3D Implantatplanung | Implantat-Bohrschablonen | Schablonengeführte Implantat-Chirurgie | Computergestützte Chirurgie
[DDC Sachgruppe]: DDC 610 / Medicine & health
Metadata
Zur LanganzeigeDOI & Zitiervorlage
Nutzen Sie bitte diesen Identifier für Zitate & Links: http://dx.doi.org/10.18725/OPARU-3984
Naziri, Eleni (2016): Klinische Genauigkeit der schablonengestützten Implantatinsertion. Open Access Repositorium der Universität Ulm und Technischen Hochschule Ulm. Dissertation. http://dx.doi.org/10.18725/OPARU-3984
Verschiedene Zitierstile >