AbstractIn this paper an impulse-radio ultra-wideband (IR-UWB) hardware demonstrator is presented, which can be used as a radar sensor for highly precise object tracking and breath-rate sensing. The hardware consists of an impulse generator integrated circuit (IC) in the transmitter and a correlator IC with an integrating baseband circuit as correlation receiver. The radiated impulse is close to a fifth Gaussian derivative impulse with (sigma) = 51 ps, efficiently using the Federal Communications Commission indoor mask. A detailed evaluation of the hardware is given. For the tracking, an impulse train is radiated by the transmitter, and the reflections of objects in front of the sensor are collected by the receiver. With the reflected signals, a continuous hardware correlation is computed by a sweeping impulse correlation. The correlation is applied to avoid sampling of the RF impulse with picosecond precision. To localize objects precisely in front of the sensor, three impulse tracking methods are compared: tracking of the maximum impulse peak, tracking of the impulse slope, and a slope-to-slope tracking of the object’s reflection and the signal of the static direct coupling between transmit and receive antenna; the slope-to-slope tracking showing the best performance. The precision of the sensor is shown by a measurement with a metal plate of 1-mm sinusoidal deviation, which is clearly resolved. Further measurements verify the use of the demonstrated principle as a breathing sensor. The breathing signals of male humans and a seven-week-old infant are presented, qualifying the IR-UWB radar principle as a useful tool for breath-rate determination.dc.description.abstract
Languageendc.language.iso
PublisherUniversität Ulmdc.publisher
ID of original publ.http://dx.doi.org/10.1109/TMTT.2013.2252185dc.relation.uri
LicenseStandarddc.rights
Link to license texthttps://oparu.uni-ulm.de/xmlui/license_v3dc.rights.uri
KeywordMedical applicationsdc.subject
KeywordMedical imagingdc.subject
KeywordUltra-wideband (UWB) circuit techniquesdc.subject
KeywordUWB communicationsdc.subject
KeywordUWB signal generation and transmissiondc.subject
Dewey Decimal GroupDDC 620 / Engineering & allied operationsdc.subject.ddc
MeSHDiagnostic imagingdc.subject.mesh
TitleIR-UWB radar demonstrator for ultra-fine movement detection and vital-sign monitoringdc.title
Resource typeWissenschaftlicher Artikeldc.type
DOIhttp://dx.doi.org/10.18725/OPARU-2508dc.identifier.doi
PPN1652932615dc.identifier.ppn
URNhttp://nbn-resolving.de/urn:nbn:de:bsz:289-vts-86158dc.identifier.urn
GNDRadarsensordc.subject.gnd
GNDUltraweitbanddc.subject.gnd
FacultyFakultät für Ingenieurwissenschaften und Informatikuulm.affiliationGeneral
Citation of original publ.IEEE transactions on microwave theory ant techniques 61 (2013), S. 2076 - 2085uulm.citationOrigPub
Date of activation2013-07-17T10:43:38Zuulm.freischaltungVTS
Peer reviewjauulm.peerReview
DCMI TypeTextuulm.typeDCMI
VTS ID8615uulm.vtsID
CategoryPublikationenuulm.category
Bibliographyuulmuulm.bibliographie


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record