An integrated readout for current sensing based on a delta sigma modulator with switched capacitor feedback
Loading...
Files
Date
2019-11-18
Authors
Rajabzadeh, Mahdi
Häberle, Matthias
Jain, Ankesh
Ortmanns, Maurits
Journal Title
Journal ISSN
Volume Title
Publication type
Document version
acceptedVersion
Distributor
Universität Ulm
Published in
ESSCIRC 2019 - IEEE 45th European Solid State Circuits Conference (ESSCIRC), 2019
Abstract
This work presents a current input continuous-time sigma-delta modulator (C-CTSDM) for biomedical application such as electrochemical impedance spectroscopy. Low noise and low area current input digitizers are required to implement high spatial resolution multi electrode arrays. In this work, the use of a switched capacitor with resistor D/A converter (SCR-DAC) for the C-CTSDM is proposed for this purpose. SCR-DACs offer the realization of large resistances on small area, which leads to a very small area consumption combined with good noise performance. A C-CTSDM with an SCR-DAC prototype is fabricated in a 180-nm CMOS process. The implemented ADC occupies an area of only 120 micro 200 micro m2. In order to increase the dynamic range (DR) of the C-CTSDM, two modes have been implemented. In
the low noise configuration an integrated in band noise of 157- pArms in the signal band width of 100 kHz with a peak SNDR of 62 dB is achieved. In the high DR configuration a peak SNDR of 72 dB with an MSA of -4.1 dB corresponding to 9.1 microA is measured. The complete C-CTSDM consumes 2 mW from a 3-
V supply and achieves a total DR of 95 dB.
Description
Faculties
Fakultät für Ingenieurwissenschaften, Informatik und Psychologie
Institutions
EU Project uulm
Other projects uulm
Citation
Is version of
Has version
Supplement to
Supplemented by
Has erratum
Erratum to
Has Part
Part of
DOI external
DOI external
10.1109/ESSCIRC.2019.8902914
Institutions
Periodical
Degree Program
DFG Project THU
EU Project THU
Other projects THU
Series
Keywords
Current readout, Direct digitization, Current-input CTSDM, SCR-DAC, Sigma delta modulator, Sigma-Delta-Wandler, Biomedizinische Technik, Lunge, Spektroskopie, Modulators (Electronics), Biomedical engineering, Electrodes, Impedance spectroscopy, Electric impedance, Lung, Spectrum analysis, DDC 610 / Medicine & health, DDC 620 / Engineering & allied operations