Objective
A prototype infrared attenuated total reflection (IR-ATR) laser spectroscopic system designed for in vivo classification of human cartilage tissue according to its histological health status during arthroscopic surgery is presented. Prior to real-world in vivo applications, this so-called osteoarthritis (OA) scanner has been tested at in vitro conditions revealing the challenges associated with complex sample matrices and the accordingly obtained sparse spectral datasets.
Methods
In vitro studies on human knee cartilage samples at different contact pressures (i.e., 0.2–0.5 MPa) allowed recording cartilage degeneration characteristic IR signatures comparable to in vivo conditions with high temporal resolution. Afterwards, the cartilage samples were assessed based on the clinically acknowledged osteoarthritis cartilage histopathology assessment (OARSI) system and correlated with the obtained sparse IR data.
Results
Amide and carbohydrate signal behavior was observed to be almost identical between the obtained sparse IR data and previously measured FTIR data used for sparse partial least squares discriminant analysis (SPLSDA) to identify the spectral regions relevant to cartilage condition. Contact pressures between 0.3 and 0.4 MPa seem to provide the best sparse IR spectra for cylindrical (d = 3 mm) probe tips.
Conclusion
Laser-irradiating IR-ATR spectroscopy is a promising analytical technique for future arthroscopic applications to differentiate healthy and osteoarthritic cartilage tissue. However, this study also revealed that the flexible connection between the laser-based analyzer and the arthroscopic ATR-probe via IR-transparent fiberoptic cables may affect the robustness of the obtained IR data and requires further improvements.
Description
Faculties
Fakultät für Naturwissenschaften
DFG Project uulm
EU Project uulm
Other projects uulm
Citation
Is version of
Has version
Supplement to
Has erratum
Erratum to
Has Part
Part of
DOI external
DOI external
10.1016/j.ocarto.2024.100466
Institutions
Periodical
Degree Program
DFG Project THU
EU Project THU
Other projects THU
Series
Collections
Endorsement
Review
Supplemented By
Referenced By
Zeitraum:
Creative Commons license
Except where otherwised noted, this item's license is described as CC BY-NC-ND 4.0 International